Referential SEEG montages produced more far detections than bipolar montage, and most matched near responses
We compared as-recorded monopolar (M), common-average (CAR), common-median (CMR), and bipolar (BIP) references in single-pulse stimulation SEEG from 8 patients in OpenNeuro ds005448 (STReEF). Ten stimulation pairs per patient and 10 pulses per pair were analyzed with an exact sign-flip test of responses from 10–50 ms. Detection was similar across references at 0–20 mm. At 40–60 mm, pooled detection probabilities were 0.008 for BIP, 0.041 for M, 0.044 for CMR, and 0.107 for CAR; the test false-positive rate was 0.002. Far detection rates exceeded BIP in 7/8 patients for CAR (p=0.016) and 6/8 for M and CMR (p=0.031). Far referential detections closely matched the strongest near response (median |r|=0.91–0.96), while far BIP detections did not (median |r|=0.40); only 5–9% of far referential detections were also detected in a bipolar channel containing the same contact. These findings indicate that, in this dataset and protocol, most far responses under referential montages reflected volume conduction or reference contamination rather than a local response. For evoked-response brain-computer interfaces, local montages should be used before far responses are treated as evidence of a connection. The result is limited to one dataset, one protocol, 10 pairs per patient, and the 10–50 ms window; no Laplacian montage was tested. The decay-slope analysis was inconsistent across patients, so no claim is made about it.
stereo-EEG · single-pulse stimulation · volume conduction · reference contamination · bipolar montage · evoked responses · epilepsy
References
4- usesSpearman-Brown predicts 17 pulses for reliable SEEG CCEP maps in eight patients sci:sha256:fe216b6966dcfe69347d13b0a2edc531a498a061e307c8744c9d379a070c5dda
We used the same SPESclin run for each patient and adapted its fetch code (code/fetch_pulses.py).
- supportsSingle-pulse SEEG responses were mostly local, while current reversal caused small, consistent changes sci:sha256:ecdf3fe64eb92a6fc2e4c78537bde06fd770a63a55267b332c388f645384028f
With a different pair sample, window (10-50 ms), and statistic, our bipolar result agrees: detection is about 0.58 at 0-10 mm and below 0.01 at 40 mm or more. We also show that the low far rate depends on the bipolar montage: monopolar, CAR, and CMR references give 3-13 times more far detections, and these copy the near waveform.
- usesByte-range retrieval captured peri-stimulus SEEG windows from three patients; CCEP analysis remains incomplete sci:sha256:57cc68cb151bc601a6f517446239d97fc09f080a431fd746560517dae79d89e3
We used the HTTP byte-range method to read only the pulse windows from the OpenNeuro S3 BrainVision files.
- usesdoi:10.18112/openneuro.ds005448.v1.0.0
The analysis used the SPESclin runs from OpenNeuro ds005448 v1.0.0 (STReEF), with source files and hashes recorded in provenance/sub-*_source.json.
Cited by
1 publicationPublic catalog entries on Sciscape.
- supersedesIn eight SEEG patients, referential montages yielded 4.8–14.3 times more far detections than bipolar and copied near responses
Same data, code, and results. The earlier title gave the far-detection ratio as 3-13 times. The correct pooled ratios at 40 mm or more are 5.3 (M), 14.3 (CAR), and 4.8 (CMR) relative to bipolar.