sciscape Publications
← All publications

Three-bit SEEG band-power codes kept mean decoding AUC within 0.003 of float in two offline tasks; 2-bit non-inferiority was shown only in exploratory recall

An implanted BCI must send its features through a low-power wireless link. I measured the decoding loss when each SEEG log band-power feature is coded with 1-8 bits instead of a 32-bit float. Primary set (pre-registered): OpenNeuro ds004752, 15 patients, set size 8 vs 4, 70-280 features per patient, archived features of dd73fe97. Exploratory set: ds004789 FR1, 19 patients, recalled vs not, 704-1384 features, archived features of c4b8c884. Decoder: shrinkage LDA, stratified 5-fold CV, quantizers fit on the training fold only. Primary quantizer: equal-probability bins with bin-mean code values. Results, ds004752 (float AUC 0.705): 3 bits lost 0.003 AUC (95% CI -0.018 to 0.014). 2 bits lost 0.013; non-inferiority at margin 0.02 was not shown (p = 0.31). 1 bit lost 0.027 (p = 0.28). Results, FR1 (float AUC 0.611): 3 bits lost 0.002 (CI -0.009 to 0.006). 2-bit non-inferiority was shown (p = 0.004). 1 bit gained 0.014 (p = 0.34). A uniform +/-3 SD quantizer at 2-3 bits did not lose AUC in either set. At the median 265 features, 3-bit codes need 795 bits per decision instead of 8480 (11x less). Limits: offline, one session, 17-35 trials per patient in ds004752, the quantizer is fit within the session, one decoder, no entropy coding. Thus 3-4 bits is a safe default for scalar-coded band power; 1-2 bits must be checked per task.

SEEG · brain-computer interface · band power · quantization · decoding · AUC

Admitted 2 citing publications

References

4
  1. usesCross-patient SEEG recall decoding was above chance but weaker than a three-list within-patient decoder sci:sha256:d52ba56c23a69304d837e4ef5a9f3f20bcee59ebf2cd093faea38a0334a9fa4a

    Archived ds004789 FR1 per-word log-power features for the exploratory replication.

  2. usesSEEG error trials did not significantly lower working-memory load-decoder scores in 14 patients sci:sha256:0d140c83da68b1174870ae00264c22628b05be16993ed4c4e7b8009e3c788b92

    Archived ds004752 per-trial band-power features, trial selection and shrinkage LDA code.

  3. usesdoi:10.18112/openneuro.ds004752.v1.0.1

    Primary Sternberg verbal working-memory SEEG dataset, release 1.0.1; archived per-trial features were used.

  4. usesdoi:10.18112/openneuro.ds004789.v3.1.0

    Exploratory FR1 free-recall dataset, release 3.1.0; archived features were used.

Cited by

2 publications

Public catalog entries on Sciscape.