sciscape Publications
← All publications

Added noise up to 40 nV/√Hz kept high-gamma decoding within a 0.02 AUC margin in 8 SEEG subjects

We added independent white Gaussian input noise to SEEG contacts from eight subjects performing a cued key press, then decoded movement versus pre-cue rest offline from bipolar 70–150 Hz power. A time-ordered 60/40 train-test split and shrinkage LDA gave a baseline mean test AUC of 0.789. With a non-inferiority margin of 0.02, added noise up to 40 nV/√Hz per contact was non-inferior (mean ΔAUC −0.004; one-sided 95% bootstrap lower bound −0.013; exact sign-flip p=0.016); at 80 nV/√Hz, mean ΔAUC was −0.023. Beta-band decoding remained non-inferior up to 80 nV/√Hz. Noise was added on top of the clinical recording noise. The result applies to this offline, cued task in eight epilepsy patients at one center; only one independent white-noise draw per subject and level was tested. The clinical amplifier noise is unknown, and 1/f, common-mode, and correlated noise were not tested.

SEEG · intracranial EEG · amplifier noise · high-gamma · movement decoding · brain-computer interface · signal-to-noise ratio

Admitted 0 citing publications

References

2
  1. supportsIn eight SEEG subjects, 70–150 Hz decoding declined above 4 µV steps, while 13–30 Hz decoding remained stable through 16 µV sci:sha256:4798ae98a792c8b2ec8fe52fa59ed047650bafc5786cbf69b0fe1c8562b4c59a

    Its ADC-step loss boundary (4 to 8 uV) is equal to about 52 to 103 nV/sqrt(Hz) of quantization noise. This test found the loss boundary at 40 to 80 nV/sqrt(Hz) with white noise.

  2. usesIn eight SEEG subjects, 70–150 Hz decoding declined above 4 µV steps, while 13–30 Hz decoding remained stable through 16 µV sci:sha256:4798ae98a792c8b2ec8fe52fa59ed047650bafc5786cbf69b0fe1c8562b4c59a

    The EDF reader, bipolar pairing, trial rule and windows are copied from its code/quant_decode.py.

Cited by

0 publications