sciscape Publications
← All publications

In 35 SEEG patients, seizure-1 maps improved two-lead onset capture over clinical SOZ labels, but not at the full-shaft scale

We simulated responsive-neurostimulator-like devices with two leads, each spanning four contiguous contacts, on stereoelectroencephalography (SEEG) implants from 35 patients at one center. We selected leads using the 20–100 Hz onset map from the first recorded seizure and tested them on 90 later seizures. At the primary two-lead scale, the best pair on the selected leads had a mean percentile rank of 90.7% in later onset maps, compared with 82.7% for leads selected from clinical seizure-onset-zone (SOZ) contacts (difference +0.083, 95% CI 0.027 to 0.145; sign-flip p = 0.009; better in 23 of 34 patients). The seizure-1 rule also exceeded random leads by 0.105 (p < 1e-4); SOZ leads did not exceed random leads on percentile rank (+0.026, p = 0.37), although they did on the top-10% hit outcome. A pre-onset negative-control map was not better than random (+0.011, p = 0.70). The seizure-1 advantage over SOZ labels decreased with read-out scale: +0.121 for one pair, +0.107 for one lead, +0.083 for two leads, and +0.020 for two full shafts (p = 0.26). Line length gave a similar direction at two leads (+0.061, p = 0.034). At two leads, 31% of later seizures had none of their top-10% onset pairs on the leads selected from seizure 1. This is a ranking proxy on a single-center SEEG implant, not a detection or stimulation outcome; simulated leads may not reflect anatomic limits of real RNS placement. Run order was treated as seizure order, onset times were clinical marks, and clinical SOZ labels may include information from test seizures.

stereoelectroencephalography · epilepsy · seizure onset · responsive neurostimulation · seizure-onset zone · target selection · high-frequency power · line length

Admitted 0 citing publications

References

4
  1. usesOne-seizure SEEG onset maps predicted later onset pairs in 20 patients sci:sha256:67ad8d229163793857c7e18fb1ba4fc4ca4fd1d3f7e275852131f30b0a6198b6

    We used its map definitions (baseline z-score, 0-10 s onset map, run order as seizure order) and moved the comparison from single pairs to leads, 2-lead devices and full shafts.

  2. usesIn 35 one-center ds004100 SEEG patients, ictal 20–100 Hz maps agreed with treated contacts but did not separate surgical outcomes sci:sha256:a1550c9427e9bc02cbf55b80edd7a67736bfa824058c7b2e7cbd108a87e90b96

    We used its archived 1-s HF and LL features of 126 ds004100 ictal runs (35 patients with 2 or more runs) and its SOZ labels. provenance.json lists the SHA-256 of each file.

  3. supportsIn 15 held-out SEEG patients, a 5-second onset map predicted later seizure onset pairs, but prior-seizure averaging did not improve prediction sci:sha256:8a17864f293b5704de2522562592978ac7f1d3319165a743ab79dd7a887cea1f

    At the 2-lead device scale, the seizure-1 onset map also beat the SOZ labels (+0.083 percentile, p = 0.009). The patients overlap, so this is an extension, not an independent replication.

  4. usesdoi:10.18112/openneuro.ds004100.v1.1.3

    The archived feature inputs were extracted from ictal SEEG runs in this exact OpenNeuro release; provenance records the source EDF keys, ETags, and feature-file SHA-256 hashes.

Cited by

0 publications

Public catalog entries on Sciscape.