A time correction increased mean agreement of later STE and MNI starts within 5 ms in 18 SEEG sleep records
We tested a time correction for released HFO detector starts. We used all 18 SEEG records in OpenNeuro ds005398 version 1.1.3. One offset per patient was chosen with early marks, then held constant for later marks. Counts stayed fixed. At 5 ms tolerance, the mean STE match fraction rose from 14.02% to 21.42%. The paired mean gain was 7.41 percentage points. The 95% patient bootstrap interval was 4.45 to 10.62 points. Fifteen patients gained; two lost; one had no change. The 2 ms and source-score checks had intervals that included zero. This result applies to released sleep marks. It does not establish neural event accuracy or clinical benefit. The package contains all source inputs, code, results, checks and repeat instructions.
high-frequency oscillations · stereoelectroencephalography · detector agreement · time correction · event timing · epilepsy
References
4- usesPast STE counts did not show a clear mean gain for next-minute Poisson MNI forecasts in 18 SEEG sleep records sci:sha256:5493db4a0b8b0c94d55724afa1f9609b3cebe4708f90aff7978253ef05b46579
This package supplied the pinned source manifest, compressed source inputs and retrieval code. All 94 inputs were checked against fresh source requests. The present correction analysis and independent checks are new.
- usesMean MNI minus STE HFO start delay was positive for every largest 20 ms match set in the first hour of 17 of 18 SEEG sleep records sci:sha256:3b7eb40fd893ce9a91e3951e7bb9e6df506351510a9a52085d1b996a2e97ebd2
The signed start-delay result in the same 18 SEEG records motivated a new early-to-later test of time correction. It did not test correction on later marks.
- usesdoi:10.18112/openneuro.ds005398.v1.1.3
This version supplied the SEEG event marks and metadata analyzed in the study.