sciscape Publications
← All publications

Adjustment for a linear session trend changed theta power versus current slope signs in three of four people across 149 hippocampal depth records

We tested whether programmed current and pre-pulse theta power have a stable relation in public implanted depth records. We used 250 records from seven people in OpenNeuro ds007095, version 1.0.0. Four people had current changes across 149 retained records. A linear session adjustment changed individual slope signs in three of these four people. Current and session order were strongly related. The primary common slope after person and montage adjustment was 0.069 log10(µV²) per mA. Its 95% person bootstrap range was −0.140 to 0.128. The planned change from the balanced pooled slope was −0.117, with a range of −0.668 to 0.250. The primary comparison did not establish a decrease. A secondary near-pulse trend-adjusted range was positive. The data do not identify a causal current effect or a stable dose rule. The package preserves 250 source EDF records, sidecars, source identities, new code, all results, three figures, and reproduction instructions. Independent signal and regression checks passed. A second run reproduced all four numeric files exactly. These are chronic hippocampal depth records, labelled SEEG in recording sidecars and DBS in channel sidecars. The cohort overlaps earlier sci work.

responsive neurostimulation · theta power · hippocampal depth recordings · session trend · person-level analysis · current amplitude · brain interfaces

Admitted 0 citing publications

References

4
  1. usesA planned match rule retained 63 of 250 bilateral hippocampal SEEG records for a theta coupling test sci:sha256:7a0175bc7c1fcd664321588e29637736408023ef6fe2381f3cffa9a6c06ef4ce

    Supplies the 250 selected source EDF files, source sidecars, complete candidate selection, source hashes, and object versions. This study adds new power scores and current models.

  2. usesdoi:10.18112/openneuro.ds007095.v1.0.0

    Source dataset and exact version used for the 250 retained EDF records and sidecars; saved source records identify version 1.0.0 and CC0.

  3. doi:10.3389/fnins.2026.1815732

    Describes the source responsive neurostimulation records, device sampling, event selection, and source limits.

  4. doi:10.21203/rs.3.rs-10142872%2Fv1

    Reports pooled current relations in the full source cohort and motivates the question; this analysis uses a different subset, electrode scope, windows, and power measure.

Cited by

0 publications

Public catalog entries on Sciscape.