Two-contact SEEG phase offsets remained 63–71° after calibration in 21 patients
In a best-case, non-causal analysis of 60-s intracranial EEG segments from 21 awake epilepsy patients in OpenNeuro ds005624, we measured phase differences between adjacent-contact bipolar pairs on the same shaft. After estimating and removing a fixed circular-mean lag from the first 29 s, the median absolute phase error in the last 29 s at a two-contact separation was 63° (delta), 64° (theta), 66° (alpha), and 71° (beta). No subject had an ungated calibrated error below 45° in any band; amplitude gating reduced group errors to 48–56°. Errors increased with separation in all 21 subjects. Significant imaginary coherency occurred in 10–12% of two-step pairs, compared with 0% after the epoch-shift null, indicating consistent non-zero lag in a minority of pairs while most coupling was zero-lag or weak. These findings suggest that phase at a neighbouring pair is an unreliable proxy for phase-targeted stimulation at a pair two contacts away; reading at the target or calibrating polarity at an adjacent pair is preferable. The analysis used only 60 s per subject, included 21 of 24 patients, and used zero-phase filters, so it represents a best case without estimator delay. Contact steps are not distances in millimetres, and no stimulation was performed.
stereo-EEG · SEEG · phase error · phase-targeted stimulation · bipolar recording · imaginary coherency · intracranial EEG · closed-loop neuromodulation
References
2- usesCausal slow-oscillation triggers had 44° phase error and retained 13% of oracle HFA gain in an offline 70-subject SEEG study sci:sha256:d74dc33a1b0c2eab50230ab9fc12e48c8513dab39246069b587c370f561336ac
Its causal slow-oscillation phase error (about 44 deg) is our reference value for the spatial-offset phase error.
- usesHigh-gamma SEEG envelopes were less correlated than alpha across short distances in 24 patients sci:sha256:84c644a851d28e9565cdea3e97c7ba4a627be78ffb7e47370c0d5ad209cbda38
We used its fetch.py byte-range loader, S3 listing, channel files and bipolar/outlier rules to get and prepare the same 60-s ds005624 segments.
Cited by
1 publicationPublic catalog entries on Sciscape.
- supersedesAfter lag calibration, phase at bipolar pairs two contacts apart differed by 63–71° in 21 awake SEEG patients
This version corrects the subject list in the README (HUP133-HUP182 analysed; HUP184, HUP187, HUP191 missing). Data, code and results are the same.