Deep SEEG contacts deviate from the anchor-bolt axis in a pediatric cohort
We measured the perpendicular distance of stereoelectroencephalography (SEEG) contact centres from the extended anchor-bolt axis in public post-implant CT trajectory derivatives for 58 pediatric patients (60 subject-sessions, 613 shafts). The deepest contact was a median 1.15 mm from the bolt axis (95% subject-bootstrap CI 1.03–1.24 mm); 21% of shaft tips were more than 2 mm away and 9.5% more than 4 mm away. Deviation increased with contact depth in all 58 subjects (mean subject median Spearman rho 0.80; one-sided sign-flip p=1e-4, the smallest possible value with 10,000 flips). This result is consistent with a shaft angle and does not establish bending. Median shaft-to-bolt angle was 1.63 degrees and median curvature was 0.37 mm. A post hoc exclusion of 18 shafts with angle >15 degrees or curvature >2 mm yielded median tip deflection 1.12 mm and 7.1% of tips above 4 mm. These data indicate that the bolt axis is not a reliable proxy for deep contact location in this cohort; post-implant imaging is needed. The public files contain no planned trajectories, so the analysis measures deviation from the bolt axis, not targeting error against a plan. The cohort is pediatric and from one center; segmentation errors may contribute to extreme values.
stereoelectroencephalography · SEEG · electrode localization · anchor bolt · trajectory · pediatric · computed tomography · brain-computer interface
References
1- usesdoi:10.18112/openneuro.ds005083.v1.0.0
Source of the SEEG trajectory derivatives analyzed in this publication (OpenNeuro ds005083 v1.0.0; exact-version metadata records CC0).
Cited by
2 publicationsPublic catalog entries on Sciscape.
- usesIn pediatric ds005083, independent SEEG CT contact-centre error was 0.246 mm 3D RMS, with 1.31-fold greater error along the shaft
We adapted its ds005083 download code and file parser. We used the same trajectory files. That work measured shaft deviation from the bolt axis. This work measures the per-contact error of the centres.
- usesZ coordinates show 0.5 mm quantization and larger error in pediatric SEEG contact centres
Same dataset and shaft geometry approach; its deviation values in mm include the z-grid error that we report.