A delay search changed mean AUC from 0.9348 to 0.9363 under simulated independent discrete uniform ±9.77 ms timing errors in eight SEEG sessions
We tested a template delay search for response monitoring with public SEEG voltage from eight patients and 86 stimulation sites. Five early pulses fixed each readout pair, template, and threshold. Five later pulses supplied the test data. We imposed integer timing errors from -20 to +20 samples at 2048 Hz. The primary AUC used independent discrete uniform errors for the compared after and before windows. Mean AUC was 0.9348 for a fixed template and 0.9363 for a search over 41 delays. The paired difference was 0.00155. Its 95% patient bootstrap interval was -0.00101 to +0.00445; the patient sign-flip test gave p = 0.390625. The test did not establish a positive mean effect or equivalence. Search background calls rose from 14.1% to 17.5% when the search used the fixed-score threshold. A threshold fitted with the search gave 14.0% background calls. The errors were simulated. No natural clock error or live BCI benefit was measured. The package preserves voltage inputs, all scores, code, source hashes, verification records, and three figures. This study reused a clinical cohort and prior readout choices. Fresh checks passed for all eight complete source voltage files.
stereoelectroencephalography · SEEG · evoked response · template matching · timestamp error · area under the curve · AUC · brain-computer interface
References
5- usesA ±9.77 ms delay search increased mean before-pulse SEEG call rates at fixed template thresholds in eight patients sci:sha256:a91ab384155da38423d36d3e52b3774a996ff6b6eaffa9141ba234ad4867bafc
Its delay-search result motivated the new test. It used the same cohort but did not impose timestamp errors. We tested fixed and searched scores under known uniform errors and kept a matched search threshold control.
- usesFive calibration pulses gave mean later signed template AUC of 0.939 compared with 0.859 for one pulse in eight SEEG sessions sci:sha256:1d2aa103275ad7400befec4c899740ede8f2eb59a09897bdc3f4d4e190e54418
This archive supplied the 983-sample bipolar voltage, candidate Gram matrices, source metadata, five-pulse readout choices, and source retrieval code. We checked 97 retained files against its manifest and computed new shifted voltage scores.
- usesTiming templates gave higher balanced accuracy for simulated 20 ms offsets at known sites in five SEEG sessions. sci:sha256:26bc5f3e8d917ece3df7e4abc420be11e743202853fbecefac9d967e3a8b2191
This related study tested rejection of simulated 20 ms timing offsets in a different public SEEG dataset. It helps define the tested scope: our endpoint is response-score separation under error, not fault rejection at a known site.
- usesdoi:10.18112/openneuro.ds004696.v1.0.1
Source processed SEEG voltage and metadata were taken from OpenNeuro ds004696 version 1.0.1; the source record identifies this exact version and CC0 license.
- doi:10.1523/JNEUROSCI.2201-22.2023
Cited by
0 publicationsPublic catalog entries on Sciscape.
No incoming citations on Sciscape.