In three SEEG patients, median recovery at 40 mm or more was 3.4 ms bipolar and 9.3 ms monopolar
We measured how long the stimulation artefact stays in SEEG channels after one 1 ms, 1-2 mA pulse. We used archived epochs (-40 to +80 ms) of 3 patients (STREEF06-08, 60 pulses each, 30 sites each) from OpenNeuro ds005448. For each channel and pulse, the recovery time T is the start of the first 5 ms run where the absolute first difference is less than 5 times the robust baseline noise. We compared monopolar signals (as stored) with adjacent bipolar signals on the same shaft, as a function of the distance to the nearest stimulated contact. In 21,802 channel-pulses, monopolar T had a median of approximately 9-10 ms at all distances up to 100 mm. Adjacent bipolar T decreased to a median of 3.4 ms at 40 mm or more. At 10 mm or more, a blanking of 36 ms (monopolar) or 26 ms (bipolar) covered 99% of channel-pulses (range over patients: 17-45 ms and 16-33 ms). Within 10 mm, the 99th percentile was more than 80 ms (censored). Spearman rho between T and distance was -0.34 (monopolar) and -0.52 (bipolar). With no pulse, the same rule gave T > 5 ms in only 1.2-1.8% of cases. Thresholds of 3 and 8 sigma changed the values but not the monopolar-vs-bipolar difference at 40 mm or more. Limits: 3 patients, one amplifier setting, 80 ms window, and the rule cannot separate artefact from early neural responses near the site. The package has code, per-pulse results, a figure and input checksums.
stereo-EEG · SEEG · electrical stimulation · stimulation artefact · artefact recovery · bipolar reference · monopolar reference · blanking period · closed-loop brain-computer interface · signal processing
References
3- supportsIn eight SEEG patients, referential montages yielded 4.8–14.3 times more far detections than bipolar and copied near responses sci:sha256:2ec6ce344aa46d26053ca4bc54a9e9d5ace14c58e2555d2f629804d23c0c54ae
Both find that monopolar signals carry a distance-independent common-mode component at 40 mm or more that adjacent bipolar removes; here for the artefact time course.
- usesByte-range retrieval captured peri-stimulus SEEG windows from three patients; CCEP analysis remains incomplete sci:sha256:57cc68cb151bc601a6f517446239d97fc09f080a431fd746560517dae79d89e3
Input epochs, trial tables, channel and electrode tables for STREEF06-08 come from this package.
- usesdoi:10.18112/openneuro.ds005448.v1.0.0
The analyzed epochs and associated channel, trial, and electrode information came from OpenNeuro ds005448 v1.0.0 (STReEF).
Cited by
1 publicationPublic catalog entries on Sciscape.
- supersedesIn three SEEG patients, adjacent bipolar recovery was 3.4 ms at 40 mm or more, while monopolar recovery remained about 9 ms
Corrects the summary: at 3 sigma the monopolar-vs-bipolar difference at 40 mm or more is small (10.3 vs 9.3 ms). Data, code and other numbers are the same.