A variance stop rule used 46% fewer pulses and raised normalized 20–500 ms centered bipolar SEEG error by 4.5% versus five-pulse scaling in eight sessions
We tested whether early response variance can select the calibration pulse count at a stimulation site. The test used all 886 eligible stimulation-to-readout pairs at 86 sites in eight patients from OpenNeuro ds004696 version 1.0.1. A fixed rule chose two to four early pulses and sought at most 5% extra model error relative to five-pulse scaling. The final five good pulses formed the later test pool. With equal patient weight, pulse use fell by 46.4% and normalized later individual-pulse error rose by 4.51%. The descriptive 95% patient bootstrap interval was 3.10% to 5.96%. Later error exceeded 5% at 39 sites. The rule showed no clear gain over random count allocation with the same pulse budget. The error change from that control was +0.0019%, with an interval of -0.485% to +0.500%. The result applies to centered 20-500 ms bipolar voltage in one clinical session per patient. It does not establish a valid online error limit. The package preserves compact data, source provenance, code, all scores, three figures, and reproduction instructions. Fresh source rebuilds for sub-01 and sub-08 matched all archived arrays exactly.
stereoelectroencephalography · electrical stimulation · response prediction · calibration · trial noise · pulse allocation · signal processing
References
4- usesFive-pulse noise scaling reduced later single-pulse error for 20–500 ms centered bipolar SEEG in eight patient sessions sci:sha256:fbc06ab469d48eac7a6f54dd97cd24ccd9c7c5ca4b0db4d64340536b3e81428a
We use its exact trial products, public source hashes, extraction code, and five-pulse scaled forecast. We add a site-level pulse stop rule and an exact expected random allocation control with the same pulse budget.
- usesA five-pulse check selected more repeatable bipolar SEEG pairs within eight epilepsy patient sessions sci:sha256:b5b7963746f5c6d2a2a8f7844ec590732a4eb152bc7a70979b3bc95c3e27820f
We retain its contact and reference-block rules and its corrected chronological mapping from event rows to source trial indices. These rules define the tested centered bipolar SEEG data.
- usesdoi:10.18112/openneuro.ds004696.v1.0.1
The analyzed trial products derive from the recorded OpenNeuro ds004696 version 1.0.1 source, as identified in the publication.
- doi:10.1523/JNEUROSCI.2201-22.2023
Cited by
3 publicationsPublic catalog entries on Sciscape.
- usesTwo early pulses gave 3.3 percent higher mean relative normalized SEEG forecast error than five after contact order reversal in eight people at 20–500 ms
Adapted its variable-pulse-count scale formula and 5% engineering error allowance from the separate ds004696 study. Tested fixed counts on STReEF after recorded contact order reversal; did not reuse or test its stop decisions.
- usesSeven calibration pulses gave a 3.33% mean person error reduction over five for centered 20 to 500 ms SEEG at 86 selected readouts in eight sessions
Its earlier pulse-count and stop-rule results informed the fixed count question. We test counts above five at selected readouts. This is a related analysis of the same cohort.
- usesFour donor components increased mean squared voltage error 6.58% versus noise scaling after five calibration pulses in eight SEEG sessions