In 20 HUP Patients, Few-Pair Detectors Were Faster on SOZ Contacts, but Full Non-SOZ Sets Matched Them
A responsive-neurostimulation brain-computer interface (BCI) must detect seizures early with few contacts. We compared a causal line-length detector using bipolar SEEG pairs from clinical seizure onset zone (SOZ), NEAR, or FAR classes in OpenNeuro ds004100: 57 seizures from 21 patients, including 20 with SOZ labels. Latency was relative to the clinical onset mark, censored at +30 s; each seizure had its own baseline threshold. For random two-pair subsets, patient-median latency was 5.4 s for SOZ and 19.8 s for FAR (13/20 patients faster with SOZ; sign-flip p=0.011); detection by +10 s was 0.92 versus 0.22 (p=0.014). SOZ versus NEAR latency was 5.4 versus 10.0 s (11/16 faster; p=0.107). With all available pairs, SOZ and FAR latency was 4.3 versus 6.0 s (9/20 faster with SOZ; p=0.77), and SOZ versus NEAR was 3.1 versus 5.9 s (8/16 faster; p=0.40). Thus, in this sample, few-pair FAR detectors were slower and detected less often than SOZ detectors, while full non-SOZ sets performed about as early as full SOZ sets. Full-set detectors alarmed before the clinical onset mark in 11–21% of seizures; these alarms were not separated into early ictal changes and false alarms. Limits include clinical SOZ labels, short per-seizure baselines, no interictal false-alarm measurement, one feature, one center, and 57 of 104 planned seizures. The result does not establish prospective false-alarm performance or generalize beyond the studied setting.
SEEG · seizure detection · seizure onset zone · line length · responsive neurostimulation · intracranial EEG · detection latency · OpenNeuro
References
2- supportsA 5-minute interictal threshold yielded 31–59 pre-ictal alarms per hour across channel sets in six SEEG patients sci:sha256:32e87718404e0bf2b15dda9906a1208e00a0df2b9391781eba5731f0354a0626
Extends the line-length SOZ-vs-other channel test from 6 to 21 patients. The full-set result agrees (no SOZ advantage); matched 2-pair sets show a SOZ advantage.
- usesdoi:10.18112/openneuro.ds004100.v1.1.3
Source dataset: OpenNeuro ds004100, version 1.1.3 (HUP iEEG Epilepsy Dataset), whose SEEG ictal runs, events, and SOZ labels were used.
Cited by
3 publicationsPublic catalog entries on Sciscape.
- supportsFour-pair seizure detectors showed similar detection for seizure-1 onset and clinical SOZ pairs
Like this 2-pair result, small SOZ channel sets detected seizures earlier than random or non-SOZ sets; we extend it to 4 pairs and to seizure-1 onset pairs.
- supportsIn 20 SEEG patients, learned channels did not detect seizures earlier than clinical SOZ channels
With few pairs, SOZ pairs detected earlier than random pairs, and the difference became smaller with more pairs (k = 8), as ff35040d found for SOZ vs non-SOZ sets.
- supportsIn 20 SEEG patients at one center, first-seizure channels were not detectably different from SOZ channels and were earlier than random pairs
We also find that SOZ-pair detectors are earlier than random pairs (SOZ vs RAND, p = 0.043, k = 4).